This report describes a program for DIBH this is relatively inexpensive to implement and has little impact on patient throughput. Multiple redundant mechanisms are incorporated to assure accurate and safe delivery of RT during DIBH. Laser alignment verifies that chest wall excursion is reliably reproduced and maintained during treatment. Chest wall excursion is also monitored independently using an infrared camera trained on a reflective marker on the chest wall. This system automatically triggers “beam off” in the event of movement of the target beyond pre-determined thresholds. Finally, physician review of cine imaging obtained during treatment provides an off-line verification of accurate RT delivery. The approach described herein lowers the investment necessary for implementation of DIBH and may facilitate broader adoption of this valuable technique.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 31, Issue 7, November 2015, Pages 733-737
Journal: Physica Medica - Volume 31, Issue 7, November 2015, Pages 733-737
نویسندگان
Bryan D. Macrie, Eric D. Donnelly, John P. Hayes, Mahesh Gopalakrishnan, Rino T. Philip, Jozef Reczek, Adam Prescott, Jonathan B. Strauss,