Humidity independent mass spectrometry for gas phase chemical analysis via ambient proton transfer reaction
ترجمه فارسی عنوان
طیف سنج جرمی مستقل رطوبت برای تجزیه شیمیایی فاز گاز از طریق واکنش انتقال پروتون محیط
کلمات کلیدی
طیف سنجی جرمی، واکنش انتقال پروتون محیط، تجزیه و تحلیل نمونه فاز گاز، رطوبت مستقل، اسپری سونوگرافی
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• A humidity independent mass spectrometric method for gas phase samples analysis.
• A universal and good sensitivity method.
• The method can real time identify plant released raw chemicals.

In this work, a humidity independent mass spectrometric method was developed for rapid analysis of gas phase chemicals. This method is based upon ambient proton transfer reaction between gas phase chemicals and charged water droplets, in a reaction chamber with nearly saturate humidity under atmospheric pressure. The humidity independent nature enables direct and rapid analysis of raw gas phase samples, avoiding time- and sample-consuming sample pretreatments in conventional mass spectrometry methods to control sample humidity. Acetone, benzene, toluene, ethylbenzene and meta-xylene were used to evaluate the analytical performance of present method. The limits of detection for benzene, toluene, ethylbenzene and meta-xylene are in the range of ∼0.1 to ∼0.3 ppbV; that of benzene is well below the present European Union permissible exposure limit for benzene vapor (5 μg m−3, ∼1.44 ppbV), with linear ranges of approximately two orders of magnitude. The majority of the homemade device contains a stainless steel tube as reaction chamber and an ultrasonic humidifier as the source of charged water droplets, which makes this cheap device easy to assemble and facile to operate. In addition, potential application of this method was illustrated by the real time identification of raw gas phase chemicals released from plants at different physiological stages.

Direct and humidity independent mass spectrometry analysis of gas phase chemicals could be achieved via ambient proton transfer ionization, ion intensity was found to be stable with humidity ranged from ∼10% to ∼100%.Figure optionsDownload as PowerPoint slide