کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1165625 | 1491070 | 2013 | 7 صفحه PDF | دانلود رایگان |

A new sensor for simultaneous determination of peroxyacetic acid and hydrogen peroxide using silver nanoparticles (Ag-NPs) as a chromogenic reagent is introduced. The silver nanoparticles have the catalytic ability for the decomposition of peroxyacetic acid and hydrogen peroxide; then the decomposition of them induces the degradation of silver nanoparticles. Hence, a remarkable change in the localized surface plasmon resonance absorbance strength could be observed. Spectra-kinetic approach and artificial neural network was applied for the simultaneous determination of peroxyacetic acid and hydrogen peroxide. Linear calibration graphs were obtained in the concentration range of (8.20 × 10−5 to 2.00 × 10−3 mol L−1) for peroxyacetic acid and (2.00 × 10−5 to 4.80 × 10−3 mol L−1) for hydrogen peroxide. The analytical performance of this sensor has been evaluated for the detection of simultaneous determination of peroxyacetic acid and hydrogen peroxide in real samples.
.Figure optionsDownload as PowerPoint slideHighlights
► A sensor for determination of peroxyacetic acid and hydrogen peroxide is introduced.
► Sol–gel silica thin film containing Ag-NPs was synthesized as a sensor membrane.
► Change in the localized surface plasmon resonance intensity of Ag-NPs was observed.
► The analytical performance of this sensor has been evaluated in disinfectant solutions.
Journal: Analytica Chimica Acta - Volume 762, 31 January 2013, Pages 87–93