کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1166210 | 1491105 | 2012 | 7 صفحه PDF | دانلود رایگان |
An extension of a capillary electrophoresis instrument coupled to a sequential injection analysis manifold was developed for automated measurements with on-line solid-phase extraction preconcentration. An in-house built capacitively coupled contactless conductivity detector was employed for sensitive detection with narrow capillaries of 25 μm internal diameter. The system was assembled into standardized 19 in. frames and racks for easy transport and mobile deployment. The system can be left running unattendedly without manual intervention with good operation stability. To demonstrate the application of the system, a method for the determination of four drugs, namely ibuprofen, diclofenac, naproxen and bezafibrate, was developed with enrichment factors of up to several hundreds. Detection of the drug residues down to the nM-scale was found possible and the method was found suitable for the detection of ibuprofen in the waste water of a hospital in Hanoi.
.Figure optionsDownload as PowerPoint slideHighlights
► Robust preconcentration is possible with a syringe-driven sequential injection based manifold.
► Separation of pharmaceutical compounds was done with capillary electrophoresis.
► Contactless conductivity detection (C4D) was used for narrow capillaries of 25 μm.
► Quantification down to the low nanomolar range was achieved.
Journal: Analytica Chimica Acta - Volume 727, 21 May 2012, Pages 1–7