کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1169419 960634 2009 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Langmuir–Blodgett film based on MEH-PPV for cholesterol biosensor
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Langmuir–Blodgett film based on MEH-PPV for cholesterol biosensor
چکیده انگلیسی

Cholesterol oxidase (ChOx) has been immobilized onto conducting poly[2-methoxy,5-(2′-ethyl-hexyloxy)-1,4-phenylene vinylene] (MEH-PPV)/stearic acid (SA) Langmuir–Blodgett film transferred onto octadecanethiol (ODT) modified gold plate. The ChOx/MEH-PPV/SA LB film bioelectrode exhibits has been characterized by FT-IR, contact angle, and atomic force microscopy. The response of the ChOx/MEH-PPV/SA LB film bioelectrode carried out using differential pulse voltammetry (DPV) studies reveal linearity from 1.29 to 12.91 mM of cholesterol concentration and response time as 30 s. This ChOx/MEH-PPV/SA bioelectrode exhibits values of correlation coefficient as 0.9939, standard deviation as 0.0029 μA and limit of detection as 1.66 mM. UV–visible spectrophotometer studies reveal that 5.2 × 10−3 U of ChOx are actively working per cm2 area of ChOx/MEH-PPV/SA LB film bioelectrode and this bioelectrode is thermally stable upto 55 °C with reusability of about 60 times.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 634, Issue 2, 23 February 2009, Pages 243–249
نویسندگان
, , , , ,