کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1171412 | 960718 | 2007 | 6 صفحه PDF | دانلود رایگان |

In this paper, a chemiluminescence (CL) micro-flow system combined with on-line solid phase extraction (SPE) is presented for determination of β-lactam antibiotics (penicillin, cefradine, cefadroxil, cefalexin) in milk. It is based on the enhancement effect of β-lactam antibiotics on the luminol-K3Fe(CN)6 CL system. The micro-flow system was fabricated from two polymethyl methacrylate (PMMA) plates (50 mm × 40 mm × 5 mm) with the microchannels of 200 μm wide and 150 μm deep. C18-modified silica gel was packed into the microchannel (length: 10 mm; width: 1 mm; depth: 500 μm) to serve as SPE device. Extraction and preconcentration of the analytes were carried out using on-line SPE micro-flow system and the selectivity of CL detection was improved. The detection limits were 0.5 μg mL−1 of penicillin, 0.04 μg mL−1 of cefradine, 0.08 μg mL−1 of cefadroxil and 0.1 μg mL−1 of cefalexin. The proposed method was also applied to analyze the β-lactam antibiotics in milk. Experimental results were in good agreement with those obtained by high performance liquid chromatography (HPLC) method with UV detection.
Journal: Analytica Chimica Acta - Volume 592, Issue 2, 5 June 2007, Pages 187–192