کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1171548 | 960724 | 2007 | 6 صفحه PDF | دانلود رایگان |

The present study reports the development and validation of a high-throughput high-performance liquid chromatographic (HPLC) assay for the determination of caffeine in food samples. The analyte was separated rapidly from sample matrix using a short monolithic column (50 mm × 4.6 mm i.d.). The flow rate was 3.0 mL min−1, while the mobile phase consisted of ACN/water (10:90, v/v). Caffeine was detected directly at 274 nm. Under the optimal HPLC conditions, the sampling rate was 60 h−1. The assay was validated for linearity, LOD and LOQ, precision, selectivity and ruggedness. The case of external calibration versus standard addition for the analysis of real samples was also examined. The proposed assay was applied to the analysis of beverages and coffee samples.
Journal: Analytica Chimica Acta - Volume 581, Issue 1, 2 January 2007, Pages 89–94