Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: The importance of frictional heating
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Practical method transfer from high performance liquid chromatography to ultra-high performance liquid chromatography: The importance of frictional heating
چکیده انگلیسی

In theory, with identical stationary phase chemistry, the transfer of an HPLC method to UHPLC conditions is straightforward and necessitates the calculation of new conditions based on column and instrument geometries. Occasionally, undesirable changes in selectivity, retention or efficiency have been reported and have been attributed to a frictional heating phenomenon that is due to the elevated generated pressure drop. In the present study, the frictional heating in a UHPLC system was evaluated experimentally under gradient elution conditions (acetonitrile/buffer at pH 3 and 9) with generated pressure drops in the range of 100–1000 bar on both 1.0 mm and 2.1 mm I.D. columns using a mixture of 10 representative basic, acidic and neutral pharmaceutical compounds. Under adiabatic conditions (i.e., still-air oven), the longitudinal temperature gradient was estimated at +4 °C, +8 °C and +16 °C at 300, 600 and 1000 bar, respectively, on a 2.1 mm I.D. column using an empirical measurement procedure. With the 1.0 mm I.D. column, these values were reduced to +3 °C, +6 °C and +12 °C, respectively. Finally, various approaches to eliminate or at least to reduce the effect of frictional heating are briefly discussed.


► The amount of frictional heating in UHPLC was empirically estimated.
► Frictional heating was measured for 1 and 2.1 mm I.D. UHPLC columns.
► The heating was equal to +4 °C, +8 °C and +16 °C at 300, 600 and 1000 bar, respectively.
► The frictional heating effect is still important with 1 mm I.D. columns.
► The best solution to compensate frictional heating is to decrease temperature in UHPLC.