Normal-phase separation of technical grade nonylphenol (t-NP, about 90% 4-nonylphenol), 4-nonylphenol mono-ethoxylate (4-NP1EO) and 4-nonylphenol di-ethoxylate (4-NP2EO) was assessed, with the inclusion of column temperature as an active variable. The compound 2,4,6-trimethylphenol was evaluated for use as internal standard. Isocratic elution with 2-propanol/hexanes mixtures from an amino-silica column and spectrometric UV detection at 277 nm were employed. Technical nonylphenol presented a significant contribution from unknown substances that eluted with retention times similar to that of 4-NP1EO. GC–MS analysis of the unknowns allowed to identify them as isomers of 2-NP. The response of the system to joint variations in flow rate, eluent composition and column temperature was investigated by means of Doehlert statistical experimental design. A model for retention of the analytes as a function of the experimental variables was proposed, and separation selectivity was studied. Selection of the optimal working zone was made through desirability function (D) calculations. Potential co-elution of 2-NP isomers with 4-NP1EO was considered when optimizing the separation. The occurrence of a restricted region of the experimental space where baseline resolution of analytes, associated impurities and internal standard results feasible (D ≠ 0) is apparent.
Journal: Journal of Chromatography A - Volume 1157, Issues 1–2, 20 July 2007, Pages 227–236