کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1212151 | 1494055 | 2015 | 6 صفحه PDF | دانلود رایگان |
• We develop a simple and quick method for the quantification of flurbiprofen in human plasma by LC-MS/MS.
• Gradient elution in LC is applied.
• Instrument conditions checked for stability prior to and after sample testing.
• A study of pharmacokinetics and bioequivalence is fulfilled.
A simple, quick and accurate LC-MS/MS method for the quantification of flurbiprofen in human plasma with indomethacin as internal standard (IS) was developed and validated. Samples were treated with methanol to precipitate proteins, then separated on a Ultimate C18 column (5 μm, 2.1 × 50 mm) with a gradient elusion process. Mobile phase A was comprised of water and formic acid, mobile phase B was comprised of acetonitrile and formic acid. Multi reaction monitoring (MRM) signals were saved on a negative ionization electrospray mass spectrometer. The calibration curve showed good linearity in the range of 40.00–10000.00 μg/L (r2 = 0.998). Intra-day RE was 0.2–2.2%. Inter-day RE was 0.5–3.4%. The samples showed good stability under the study conditions. No significant matrix effect was observed. The established method was then applied to a bioequivalence study of a flurbiprofen axetil formulation.
Journal: Journal of Chromatography B - Volumes 993–994, 1 July 2015, Pages 69–74