کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1213144 | 1494112 | 2013 | 8 صفحه PDF | دانلود رایگان |

Response surface methodology (RSM) was utilized for rapid and systematic optimization of on-line solid-phase extraction (SPE) parameters to maximize the response and separation of WM-5. The optimization was performed with Box–Behnken designs. Four major parameters were investigated for their contributions to the response and separation of WM-5, with a total of 29 experiments being performed for each instrument, respectively. Quantitative determination of WM-5 in mouse plasma was performed to evaluate the statistical significance of the parameters on chromatographic response. A fully automated on-line SPE and high-performance liquid chromatography (HPLC) with diode array detection (DAD) method was developed for the determination of WM-5 in mouse plasma. Calibration curve with good linearity (r = 0.9989) was obtained in the range of 20–4000 ng/mL in mouse plasma. The limit of detection (LOD) and lower limit of quantification (LLOQ) of the assay were 6 ng/mL and 20 ng/mL, respectively. The overall intra-day and the inter-day variations were less than 1.90%. The recovery of the method was in the range of 93.74–96.33% with RSD less than 3.06%. The optimized method demonstrated good performance in terms of specificity, LLOQ, linearity, recovery, precision and accuracy, and was successfully applied to quantify WM-5 in mouse plasma to support the pharmacokinetic study.
► Fully automated on-line SPE–HPLC utilized for PK evaluation of WM-5.
► Response surface methodology used for optimization of on-line SPE parameters.
► On-line SPE removed the time-consuming, tedious and costly manual process.
► The method is simple, fast, specific and sensitive.
Journal: Journal of Chromatography B - Volumes 923–924, 1 April 2013, Pages 8–15