کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1215208 966967 2008 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Validation of an HPLC/MS/MS method with isotopic dilution for quantitative determination of trans,trans-muconic acid in urine samples of workers exposed to low benzene concentrations
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Validation of an HPLC/MS/MS method with isotopic dilution for quantitative determination of trans,trans-muconic acid in urine samples of workers exposed to low benzene concentrations
چکیده انگلیسی

Urinary trans,trans-muconic acid (t,t-MA), a biomarker of benzene exposure, is usually determined by HPLC methods with detection by either UV or, more recently, electrospray tandem mass spectrometry. However, not all these methods have been fully validated for quantitative analysis. This paper presents an HPLC/MS/MS method for reliable quantitative determination of t,t-MA that uses a commercial deuterium-labeled isotope as internal standard; the matrix effect has been evaluated and LOD is 0.22 μg/L. We used this method to test 200 urine samples, 175 of them collected at end-of-shift from workers in an oil refinery.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography B - Volume 867, Issue 1, 1 May 2008, Pages 26–31