Capillary isoelectric focusing method development and validation for investigation of recombinant therapeutic monoclonal antibody
ترجمه فارسی عنوان
ایجاد و اعتبار سنجی روش تمرکز فیزیکی کپسولی ایزوالکتریک برای بررسی آنتی بادی منوکلونال درمان نوترکیب
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• A cIEF method was developed to investigate antibodies in the pI range 7–9 with optimal resolution.
• No urea or salt was used for sample mixture.
• A validation procedure was performed according to ICH Q2(R1).
• RSD% < 0.2 and 4.0 for isoelectric points and migration times.

Capillary isoelectric focusing (cIEF) is a basic and highly accurate routine analytical tool to prove identity of protein drugs in quality control (QC) and release tests in biopharmaceutical industries. However there are some “out-of-the-box” applications commercially available which provide easy and rapid isoelectric focusing solutions for investigating monoclonal antibody drug proteins. However use of these kits in routine testings requires high costs.A capillary isoelectric focusing method was developed and validated for identification testing of monoclonal antibody drug products with isoelectric point between 7.0 and 9.0. A method was developed providing good pH gradient for internal calibration (R2 > 0.99) and good resolution between all of the isoform peaks (R = 2), minimizing the time and complexity of sample preparation (no urea or salt used). The method is highly reproducible and it is suitable for validation and method transfer to any QC laboratories. Another advantage of the method is that it operates with commercially available chemicals which can be purchased from any suppliers. The interaction with capillary walls (avoid precipitation and adsorption as far as possible) was minimized and synthetic isoelectric small molecular markers were used instead of peptide or protein based markers.The developed method was validated according to the recent ICH guideline (Q2(R1)). Relative standard deviation results were below 0.2% for isoelectric points and below 4% according to the normalized migration times. The method is robust to buffer components with different lot numbers and neutral capillaries with different type of inner coatings. The fluoro-carbon coated column was chosen because of costs-effectivity aspects.

Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 114, 10 October 2015, Pages 53–61
نویسندگان
, , ,