Development and validation of a reversed-phase HPLC method for the simultaneous analysis of simvastatin and tocotrienols in combined dosage forms
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Development and validation of a reversed-phase HPLC method for the simultaneous analysis of simvastatin and tocotrienols in combined dosage forms
چکیده انگلیسی

A RP-HPLC method for the simultaneous analysis of tocotrienol isoforms (TRF) and simvastatin (SIM) in SIM–TRF nanoparticles (NPs) was developed. Analytes were monitored by UV detection at 238 and 295 nm for SIM and TRF, respectively, using a gradient methanol/water elution. Calibration curves for TRF and SIM were linear over concentration range of 20–80 μg/mL and 1–10 μg/mL with correlation coefficients 0.9990 and 0.9991, respectively. The recovery of TRF and SIM from the NPs was in the range from 97.35 to 102.19% and from 92.71 to 104.35%, respectively. This developed method was successfully employed in quantifying both drugs in NPs for future use in cancer therapy.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 49, Issue 4, 1 May 2009, Pages 950–956