کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1224409 967921 2007 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Development and full validation of a sensitive quantitative assay for the determination of clemastine in human plasma by liquid chromatography-tandem mass spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Development and full validation of a sensitive quantitative assay for the determination of clemastine in human plasma by liquid chromatography-tandem mass spectrometry
چکیده انگلیسی
A sensitive high-performance liquid chromatography-tandem mass spectrometry (LC/MS/MS) method was developed and validated for the determination of clemastine in human plasma. After having been extracted from plasma samples by ethyl acetate, clemastine and internal standard, diphenhydramine, were separated on a C18 column. Detection was performed on Thermo Finnigan TSQ Quantum triple quadrupole mass spectrometer by selected reaction monitoring (SRM) mode via electrospray ionization (ESI) source. The method was linear in the concentration range of 5.0-1000.0 pg/ml for clemastine. The intra- and inter-day precisions were within 13.4% and the deviations were between −1.1% and 5.6%. The fully validated LC/ESI-MS/MS method has been successfully applied to the preliminary pharmacokinetic study in healthy male Chinese volunteers.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 44, Issue 4, 15 August 2007, Pages 924-930
نویسندگان
, , , , , ,