کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1225050 967944 2006 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Determination of ruthenium in pharmaceutical compounds by graphite furnace atomic absorption spectroscopy
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of ruthenium in pharmaceutical compounds by graphite furnace atomic absorption spectroscopy
چکیده انگلیسی

A graphite furnace atomic absorption (GFAA) spectrometric method for the determination of ruthenium (Rh) in solid and liquid pharmaceutical compounds has been developed. Samples are dissolved or diluted in dimethyl sulfoxide (DMSO) without any other treatment before they were analyzed by GFAA with a carefully designed heating program to avoid pre-atomization signal loss and to achieve suitable sensitivity. Various inorganic and organic solvents were tested and compared and DMSO was found to be the most suitable. In addition, ruthenium was found to be stable in DMSO for at least 5 days. Spike recoveries ranged from 81 to 100% and the limit of quantitation (LOQ) was determined to be 0.5 μg g−1 for solid samples or 0.005 μg ml−1 for liquid samples based a 100-fold dilution. The same set of samples was also analyzed by ICP-MS with a different sample preparation method, and excellent agreement was achieved.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 41, Issue 1, 11 April 2006, Pages 43–47
نویسندگان
, , , , ,