کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1239942 1495722 2012 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Novel application of the electrolyte cathode atmospheric glow discharge: Atomic absorption spectrometry studies
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Novel application of the electrolyte cathode atmospheric glow discharge: Atomic absorption spectrometry studies
چکیده انگلیسی

The analytical characteristics of a capillary design of the electrolyte cathode atmospheric glow discharge (ELCAD), operated with a W-rod anode at a discharge current of 70 mA and a discharge voltage of 950 V, were exploited through spatially resolved atomic absorption spectrometry (AAS) experiments. For this purpose, the ELCAD cell, placed on a platform adjustable with micrometer screws, was inserted into the optical path of a commercial line-source AAS instrument. A flow injection system was developed and applied to introduce 3 mL of aqueous standards of a set of environmentally relevant metals (Ca, Cd, Cr, Cu, Na, Pb and Zn) into the plasma. The analyte atom distribution along the vertical axis of the conically-shaped ELCAD plasma (height: 3.5 mm) is element specific. All the absorbance maxima are observed in the near cathode region (e.g., in the range of 0.5–1.0 mm from the cathode), while the AA signal smoothly fades towards the anode. Several spectrochemical buffers (citric acid, EDTA, chlorides of Ca, Cs, La, Li, and Na) were studied for improving the sensitivity of the AAS determinations for Cr. A significant increase in the sensitivity (20%) was found only with the addition of 0.55% (m/v) La solution. The limit of detection data for Cd, Cu, Na and Zn are 3.4, 4.2, 9.2 and 0.9 mg L− 1, respectively. The AAS calibration curves for Cd, Cu, Na and Zn are linear up to 75, 200, 100 and 25 mg L− 1, respectively.


► A new analytical monitoring technique by combining ELCAD with AAS is developed.
► The vertical distribution of various analyte atoms in ELCAD plasma is recorded.
► Spectrochemical agents are tested to increase the sensitivity of ELCAD-AAS method.
► Analytical performance is reported for some environmentally relevant elements.
► The analytical performance is compared to those found in classic AAS/AES methods.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 77, November 2012, Pages 52–57
نویسندگان
, , , ,