کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1240024 1495733 2011 7 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Measurement of gas kinetic temperatures for polyatomic ions in inductively coupled plasma-mass spectrometry: Validation and refinements
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Measurement of gas kinetic temperatures for polyatomic ions in inductively coupled plasma-mass spectrometry: Validation and refinements
چکیده انگلیسی

The general method of comparing measured ion ratios to calculated ion ratios to determine a gas kinetic temperature (Tgas) is reviewed. Various mathematical refinements to the calculated partition functions are examined for their effect on the determined Tgas. It is found that (a) excited electronic states should be included for ArO+, neutral NO, and O2; (b) a 10% error in solvent load, sample gas flow rate, vibrational constant (ω), rotational constant (B) or measured ion ratio produces only a 1 to 3% error in Tgas; (c) a 10% error in dissociation energy (D0) creates nearly a 10% error in Tgas; and (d) high temperature corrections to the partition functions produce minimal change and can generally be neglected.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 66, Issues 3–4, March–April 2011, Pages 248–254