کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1241181 969172 2008 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Determination of Se, Pb, and Sb by atomic fluorescence spectrometry using a new flameless, dielectric barrier discharge atomizer
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of Se, Pb, and Sb by atomic fluorescence spectrometry using a new flameless, dielectric barrier discharge atomizer
چکیده انگلیسی

A flameless atomizer for atomic fluorescence spectrometry (AFS), based on an atmospheric pressure dielectric barrier discharge, has been developed for the atomization of hydride-forming elements, such as Se, Sb and Pb. The atomizer (8 mm o.d, 35 mm length) was operated at a power less than 50 W. The discharge was sustained with argon at the flow rate of 0.85 L min− 1 after optimization. The characteristics of the atomizer and the effects of different parameters (power, gas flow rate, and KBH4 concentration) are investigated. The most attractive feature of this atomizer is its low operation temperature (~ 52 °C, detected at the outlet of the atomizer by a thermocouple), allowing both the radiation source and the detector to be placed in close proximity with the atomizer. The analytical performance of the atomizer has been evaluated, and detection limits for Se, Sb and Pb obtained with the present technique were 0.08, 0.11 and 0.27 μg L− 1, respectively. The accuracy of the system was verified by the determination of Se, Sb and Pb in reference material of spinage GBW 10015. The concentrations of Se, Sb, and Pb determined by the present technique agreed well with the reference values (Se: 92 ± 24 mg kg− 1, Sb: 43 ± 14 mg kg− 1, Pb: 11.1 ± 0.9 mg kg− 1). This detector is very promising for field elements detection with portable AFS.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Spectrochimica Acta Part B: Atomic Spectroscopy - Volume 63, Issue 3, March 2008, Pages 431–436
نویسندگان
, , , , ,