Microfluidic system for high-throughput immunoglobulin-E analysis from clinical serum samples
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Microfluidic system for high-throughput immunoglobulin-E analysis from clinical serum samples
چکیده انگلیسی


• A microfluidic system is capable of high-throughput analyses of sIgE from serum.
• Sample consumption sharply decreased to <0.05 µL by this system.
• Good co-relationship (r=0.92) was found between this system and the UniCAP system.
• The detection limit of this system was lower than that obtained by UniCAP system.

Rapid and high-throughput analytical techniques for IgE that requires a small serum amount are very important, especially for pediatric patients. In these patients, blood is collected from veins, which is painful compared to fingertip blood collection. Herein, a novel microfluidic system capable of high-throughput parallel analyses of allergen-specific IgE from small amounts of patient serum was successfully developed. A six-plex immunoassay was constructed within a microfluidic chip, and the entire system was validated using samples from clinical patients. Major antigens from house dust mite (Dermatophagoides farinae and Blomia tropicalis), cat (Felis domesticus), fungus (Cladosporium herbarum), ragweed (Humulus japonicas), and tree pollen (Platanus acerifolia) were used as analysis targets. Sample consumption decreased to <0.05 µL compared with the 480 µL serum consumption by fluoroenzyme immunoassay (UniCAP system Pharmacia Diagnostics AB, Uppsala, Sweden), the 50 µL serum consumption by enzyme-linked immune sorbent assay (ELISA), or the 1.5 µL serum consumption by conventional protein chip analysis. Analysis duration, reagent cost, and total cost for each measurement were also considerably decreased. The assay showed good accuracy and sensitivity toward the clinical samples. A significant correlation of allergen-specific IgE levels was found among the microfluidic assay, UniCAP system, and ELISA.

Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 143, 1 October 2015, Pages 83–89
نویسندگان
, , , , , , , , ,