Quantitative analysis of estradiol and six other steroid hormones in human saliva using a high throughput liquid chromatography–tandem mass spectrometry assay
ترجمه فارسی عنوان
تجزیه و تحلیل کمی از استروییدول و شش هورمون دیگر استروئید در بزاق انسان با استفاده از روش کروماتوگرافی مایع بالا
کلمات کلیدی
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
چکیده انگلیسی


• A method can measure estradiol and six other steroid hormones in human saliva.
• An on-line SPE method is used to shorten sample preparation times.
• The method has high sensitivity, high selectivity and high throughput.
• Seven steroid hormones could be reliably detected at physiological concentrations.

The aim of the current research was to develop a fast and sensitive analytical strategy that allows for the simultaneous measurement of estradiol and other steroid hormones in human saliva. For this purpose, we established an assay using high performance liquid chromatography–tandem mass spectrometry (LC–MS/MS) with Atmospheric Pressure Chemical Ionization (APCI) coupled with on-line solid phase extraction (SPE). The protocol was designed for the simultaneous identification of estradiol, cortisol, cortisone, testosterone, progesterone, corticosterone and dehydroepiandrosterone (DHEA) from samples of 100 µL saliva. After protein precipitation, the sample was injected into the LC–MS/MS system for direct measurement. The protocol involved minimal sample preparation and could be run with throughput times of 5.20 min.Results indicated linearity of the method for all steroid hormones over ranges of 0.001–10 ng/mL (0.01–20 ng/mL for DHEA) with linear correlation coefficients of r=0.999 for each steroid. Intra- and inter-assay coefficients of variance were between 4.3% and 10.8%. The lower limits of quantification (LOQ) were below (or equal to) 5 pg/mL for all steroids, except for DHEA for which the LOQ was 10 pg/mL. A re-analysis of 16 saliva samples that had produced unusual estradiol values when quantified by immunoassay showed normal, endogenous concentrations for all samples when measured by the current method.In conclusion, we describe an LC–MS/MS method which can be routinely employed in physiological and psychological research laboratories allowing high analytical specificity and sensitivity despite minimal sample processing and short throughput times.

Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 143, 1 October 2015, Pages 353–358
نویسندگان
, , ,