A flow-based enzyme-linked immunosorbent assay on a polydimethylsiloxane microchip for the rapid determination of immunoglobulin A
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
A flow-based enzyme-linked immunosorbent assay on a polydimethylsiloxane microchip for the rapid determination of immunoglobulin A
چکیده انگلیسی

A flow-based enzyme-linked immunosorbent assay (ELISA) on a polydimethylsiloxane (PDMS) microchip has been developed for the rapid determination of immunoglobulin A (IgA). The analytical principle of this integrated method is the same as the conventional sandwich-type ELISA. A primary antibody (anti-IgA) was adsorbed on the surface of a PDMS microchannel, and then an antigen (IgA) and a secondary antibody (anti-IgA HRP labeled) were reacted successively. The resulting antigen–antibody complex, fixed on the surface of the microchannel, was detected using Amplex® Red and a fluorescent imaging system. The calibration curve of the IgA standard solution was linear in the range of 0–50 ng/mL at the flow rate of 10 μL/min. This flow rate corresponds to the reaction time of 4.8 s. Compared to the conventional assay on a 96-well microtiter plate, the present assay on the microchip dramatically shortened the reaction time necessary for the enzyme–substrate reaction from 30 min to 4.8 s, i.e., to 1/375. The amounts of the reagent and sample were also reduced to 1/100 compared to the 96-well microtiter plate.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 70, Issue 1, 15 August 2006, Pages 122–127