کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1245300 969718 2008 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Laser-induced breakdown spectroscopy for simultaneous determination of Sm, Eu and Gd in aqueous solution
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Laser-induced breakdown spectroscopy for simultaneous determination of Sm, Eu and Gd in aqueous solution
چکیده انگلیسی

This paper reports studies on the determination of trace levels of samarium, europium and gadolinium in aqueous samples by laser-induced breakdown spectroscopy (LIBS). In this work, a membrane-based filter paper was used as a sample support for the liquid samples. The laser-induced plasma was produced in air at atmospheric pressure, using a pulsed Nd:YAG laser. Calibration standards and synthetic mixtures of these lanthanides were prepared using solutions prepared from respective high purity oxides. Linear calibration was obtained for Sm, Eu and Gd by normalizing the intensities of lanthanides emission lines with respective to C(I) 193.029 nm emission line. The concentrations of Sm, Eu and Gd were then determined in a solution containing a mixture of these lanthanides. The concentrations of individual lanthanides were obtained within 5% of the expected values. Limits of detection were found to be 1.3 ppmw (Sm), 1.9 ppmw (Eu) and 2.3 ppmw (Gd).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 77, Issue 1, 19 October 2008, Pages 256–261