کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1245343 1495854 2006 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Validation of mercury determination by solid sampling Zeeman atomic absorption spectrometry and a specially designed furnace
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Validation of mercury determination by solid sampling Zeeman atomic absorption spectrometry and a specially designed furnace
چکیده انگلیسی

Certified reference materials (CRMs) of different origin were used to validate the direct determination of total mercury by solid sampling Zeeman atomic absorption spectrometry (SS-ZAAS) and a specially designed furnace. The temperature program provides only for one step. Atomisation of mercury and pyrolysis of the matrix is performed at a constant temperature in the range of 900–1000 °C. Calibration points achieved by CRMs and aqueous solutions are covered by one calibration line, indicating the absence of matrix effects. Relatively high amounts of chlorine, known for causing problems in mercury determination do not influence analytical results. The excellent accuracy of the method results in a very good agreement with the certified values. The precision of SS-ZAAS measurements in a range from 0.5 to 50 ng Hg does not exceed 3% R.S.D. A limit of quantification of 0.008 μg g−1 Hg was achieved.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 70, Issue 5, 15 December 2006, Pages 962–965
نویسندگان
, ,