کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1421370 | 986407 | 2012 | 9 صفحه PDF | دانلود رایگان |

ObjectiveTo investigate the influence of implant design factors in terms of bone integrity and implant stability.Materials and methodsA 3D parametric CAD model was developed. Then, once domain settings and boundary conditions were defined, a 3D FEM model was created. To simulate the physical interaction at the bone–implant interface, identity pairs were introduced. After generating different design scenarios with a DOE approach, the most significant design factors were obtained.ResultsThis study showed that the geometry of the screw thread highly influenced the implant stability. In particular the degree of bone damage became minimal when adopting 0.40 mm for the thread width and 0.05 mm for the thickness.SignificanceThread width and thickness play a crucial role to reduce induced stresses and damage in bone. Considering these preliminary results, future improvements should focus on investigating also two-factor and higher interactions to better understand the implant loading mechanism.
Journal: Dental Materials - Volume 28, Issue 8, August 2012, Pages 919–927