Factors Affecting the Arterial Distension Waveform Derived from Tissue Doppler Imaging (TDI): An In Vitro Study on Precision
دانلود مقاله + سفارش ترجمه
دانلود مقاله ISI انگلیسی
رایگان برای ایرانیان
موضوعات مرتبط
مهندسی و علوم پایه
فیزیک و نجوم
آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله

چکیده انگلیسی
An investigation was made into the effect of acquisition parameters on the distension waveform estimated from tissue Doppler imaging (TDI). Physiological distension waveforms were generated using a compliant wall phantom. Distensions derived over a range of scanning geometries and transducer pressures were compared with those obtained in optimised scanning conditions. The estimated maximum distension decreased with scanning depth by 7% between 20 mm and 44 mm below the phantom surface, with an increase in transducer-vessel angle (by 22% from 0° to 24°) and with a decrease in scan plane-vessel coincidence (by 34% from coincidence to 2 mm from the vessel central axis). An increase with transducer pressure was observed (by 20% from contact to high exerted pressure). (E-mail: P.Hoskins@ed.ac.uk)