• We have compared left breast TomoDirect treatments with breath-hold techniques.
• 60 treatment plans were recalculated for 15 patients.
• TomoDirect and breath-hold 3D conformal radiotherapy are complementary techniques.
Background and purposeTomoDirect (TD) can only operate in free-breathing. The purpose of this study is to compare TD with breath-hold 3D conformal radiotherapy (3DCRT) and intensity modulated radiotherapy (IMRT) techniques for left breast treatments, and to determine if the lack of respiratory gating is a handicap for cardiac sparing.Materials and methods15 patients treated for left breast had two computed tomography simulation, in free breathing (FB) and in deep-inspiration breath-hold (DIBH). Four treatments were planned: TD-FB, 3DCRT-FB, 3DCRT-DIBH and IMRT-DIBH. Dose to PTV, heart, lungs, right breast and patient were compared.ResultsA slightly lower cardiac mean dose is found for 3DCRT-DIBH than for TD-FB group (1.99 Gy Vs 2.89 Gy, p = 0.0462), while no statistical difference is found for heart V20. TD-FB plans show the best PTV dose homogeneity (0.053, p < 0.001) and the lowest left lung mean dose (5.16 Gy, p < 0.001). No major differences are found for the other organs.ConclusionsTomoDirect and breath-hold 3DCRT are complementary techniques for left breast treatments: for a minority of patients, respiratory gating is mandatory to lower cardiac dose; for the remaining majority of patients, TomoDirect achieves better PTV homogeneity and reduced left lung dose, with cardiac dose equivalent to 3DCRT-DIBH.