Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens
موضوعات مرتبط
علوم زیستی و بیوفناوری ایمنی شناسی و میکروب شناسی میکروب شناسی
پیش نمایش صفحه اول مقاله
Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens
چکیده انگلیسی

The Abbott Real-Time HIV-1 assay was evaluated for its performance in quantification of human immunodeficiency virus type 1 (HIV-1) RNA in dried blood spot (DBS) samples. In total, 169 blood samples with detectable plasma HIV-1 RNA were used to extract RNA from paired DBS and liquid plasma samples, using the automated Abbott m Sample Preparation System (m2000sp). HIV-1 RNA was then quantitated by the m2000rt RealTime analyser. RNA samples suitable for real-time PCR were obtained from all but one (99.4%) of the DBS samples and HIV-1 RNA was detected in 163/168 (97.0%) samples. The correlation between HIV-1 RNA values measured in paired DBS and plasma samples was very high (r = 0.882), with 78.5% and 99.4% of cases differing by <0.5 and 1.0 log, respectively. Retesting of DBS replicates following 6 months of storage at 2–8°C showed no loss of HIV-1 RNA in a subset of 89 samples. The feasibility of DBS testing coupled with automated sample processing, and the use of a latest-generation FDA-approved real-time PCR-based system, represents an encouraging first step for viral load measurement in reference centres in developing countries where access to antiretroviral therapy is expanding.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Microbiology and Infection - Volume 15, Issue 1, January 2009, Pages 93–97
نویسندگان
, , , , , , , , , ,