Contribution of geometric design parameters to knee implant performance: Conflicting impact of conformity on kinematics and contact mechanics
ترجمه فارسی عنوان
نقش پارامترهای طراحی هندسی در عملکرد ایمپلنت زانو: اثرات متضاد سازگاری در سینماتیک و مکانیک تماس
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی ارتوپدی، پزشکی ورزشی و توانبخشی
چکیده انگلیسی


• This study quantified the causal relationship between implant geometry and its function.
• Distal radii of femur and tibia affected contact pressure and A-P ROM of implant.
• Posterior radii of femur and tibia influenced contact pressure and I-E ROM of implant.
• Femoral frontal radius influenced contact pressure and I-E ROM of implant.
• Such investigations can be used to potentially enhance the future knee implant design.

BackgroundArticular geometry of knee implant has a competing impact on kinematics and contact mechanics of total knee arthroplasty (TKA) such that geometry with lower contact pressure will impose more constraints on knee kinematics. The geometric parameters that may cause this competing effect have not been well understood. This study aimed to quantify the underlying relationships between implant geometry as input and its performance metrics as output.MethodsParametric dimensions of a fixed-bearing cruciate retaining implant were randomized to generate a number of perturbed implant geometries. Performance metrics (i.e., maximum contact pressure, anterior–posterior range of motion [A-P ROM] and internal–external range of motion [I-E ROM]) of each randomized design were calculated using finite element analysis. The relative contributions of individual geometric variables to the performance metrics were then determined in terms of sensitivity indices (SI).ResultsThe femoral and tibial distal or posterior radii and femoral frontal radius are the key parameters. In the sagittal plane, distal curvature of the femoral and tibial influenced both contact pressure, i.e., SI = 0.57; SI = 0.65, and A-P ROM, i.e., SI = 0.58; SI = 0.6, respectively. However, posterior curvature of the femoral and tibial implants had a smaller impact on the contact pressure, i.e., SI = 0.31; SI = 0.23 and a higher impact on the I-E ROM, i.e., SI = 0.72; SI = 0.58. It is noteworthy that in the frontal plane, frontal radius of the femoral implant impacted both contact pressure (SI = 0.38) and I-E ROM (SI = 0.35).ConclusionFindings of this study highlighted how changes in the conformity of the femoral and tibial can impact the performance metrics.