A systematic evaluation of three different cardiac T2-mapping sequences at 1.5 and 3T in healthy volunteers
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
A systematic evaluation of three different cardiac T2-mapping sequences at 1.5 and 3T in healthy volunteers
چکیده انگلیسی


• Three different T2-mapping sequences are compared at 1.5 and 3T.
• T2 values vary significantly between field strengths and MR sequences.
• Segmental reference T2 values for each sequence at 1.5 and 3T are presented.
• Phantom experiments reveal a T1 shine through effect for the T2prep sequence.

BackgroundPrevious studies showed that myocardial T2 relaxation times measured by cardiac T2-mapping vary significantly depending on sequence and field strength. Therefore, a systematic comparison of different T2-mapping sequences and the establishment of dedicated T2 reference values is mandatory for diagnostic decision-making.MethodsPhantom experiments using gel probes with a range of different T1 and T2 times were performed on a clinical 1.5T and 3T scanner. In addition, 30 healthy volunteers were examined at 1.5 and 3T in immediate succession. In each examination, three different T2-mapping sequences were performed at three short-axis slices: Multi Echo Spin Echo (MESE), T2-prepared balanced SSFP (T2prep), and Gradient Spin Echo with and without fat saturation (GraSEFS/GraSE). Segmented T2-Maps were generated according to the AHA 16-segment model and statistical analysis was performed.ResultsSignificant intra-individual differences between mean T2 times were observed for all sequences. In general, T2prep resulted in lowest and GraSE in highest T2 times. A significant variation with field strength was observed for mean T2 in phantom as well as in vivo, with higher T2 values at 1.5T compared to 3T, regardless of the sequence used. Segmental T2 values for each sequence at 1.5 and 3T are presented.ConclusionsDespite a careful selection of sequence parameters and volunteers, significant variations of the measured T2 values were observed between field strengths, MR sequences and myocardial segments. Therefore, we present segmental T2 values for each sequence at 1.5 and 3T with the inherent potential to serve as reference values for future studies.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Radiology - Volume 84, Issue 11, November 2015, Pages 2161–2170
نویسندگان
, , , , , ,