PurposeTo compare the reproducibility and accuracy of abdominal aortic aneurysm (AAA) maximal diameter (D-max) measurements using segmentation software, with manual measurement on double-oblique MPR as a reference standard.Materials and methodsThe local Ethics Committee approved this study and waived informed consent. Forty patients (33 men, 7 women; mean age, 72 years, range, 49–86 years) had previously undergone two CT angiography (CTA) studies within 16 ± 8 months for follow-up of AAA ≥35 mm without previous treatment. The 80 studies were segmented twice using the software to calculate reproducibility of automatic D-max calculation on 3D models. Three radiologists reviewed the 80 studies and manually measured D-max on double-oblique MPR projections. Intra-observer and inter-observer reproducibility were calculated by intraclass correlation coefficient (ICC). Systematic errors were evaluated by linear regression and Bland–Altman analyses. Differences in D-max growth were analyzed with a paired Student's t-test.ResultsThe ICC for intra-observer reproducibility of D-max measurement was 0.992 (≥0.987) for the software and 0.985 (≥0.974) and 0.969 (≥0.948) for two radiologists. Inter-observer reproducibility was 0.979 (0.954–0.984) for the three radiologists. Mean absolute difference between semi-automated and manual D-max measurements was estimated at 1.1 ± 0.9 mm and never exceeded 5 mm.ConclusionSemi-automated software measurement of AAA D-max is reproducible, accurate, and requires minimal operator intervention.
Journal: European Journal of Radiology - Volume 77, Issue 3, March 2011, Pages 502–508