کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
4228795 | 1609864 | 2006 | 5 صفحه PDF | دانلود رایگان |

ObjectiveTo evaluate fat-suppressed (FS) proton-density-weighted (PDw) turbo spin-echo (TSE) magnetic resonance imaging for the detection of anterior and posterior cruciate ligament lesions in comparison to arthroscopy.Materials and methodsIn a prospective study 31 knee joints were imaged on a 1.5 T MR scanner (Vision®, Siemens, Erlangen) prior to arthroscopy using following sequences: (a) sagittal FS-PDw/T2w TSE (TR/TE: 4009/15/105 ms); (b) sagittal PDw/T2w TSE (TR/TE:3800/15/105 ms). Further imaging parameters: slice thickness 3 mm, FOV 160 mm, matrix 256 × 256. A total of 62 anterior and posterior cruciate ligaments (ACL/PCL) were evaluated, standard of reference was arthroscopy. Sensitivity, specificity, positive (ppv) and negative predictive value (npv) and accuracy were calculated.ResultsTwenty-one cruciate ligament ruptures were detected in arthroscopy, 19 ACL- and 2 PCL-ruptures (on MRI 34/124, 25/62 ACL, 9/62 PCL lesions). For all four sequences in the 31 patients with arthroscopic correlation sensitivity, specificity, ppv, npv and accuracy were 86%, 98%, 95%, 93% and 94% for detection of tears, and 84%, 100%, 100%, 80% and 90% for ACL-ruptures respectively. The two PCL-ruptures were true positive in all sequences, one intact PCL was diagnosed as torn (false positive).ConclusionsFat-suppressed PDw/T2w TSE-MR sequences are comparable to PDw TSE sequences for the detection of ACL/PCL-lesions.
Journal: European Journal of Radiology - Volume 58, Issue 3, June 2006, Pages 411–415