In Vitro Evaluation of the Insertion Force for Different Dilator Tip Configurations
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
In Vitro Evaluation of the Insertion Force for Different Dilator Tip Configurations
چکیده انگلیسی
The purpose of this study was to evaluate the tissue penetration insertion force of different dilator tip geometries in simulated tissue. Four different dilator tip designs-conical (control), triangular, diamond, and biconvex-were appraised. The penetration force for each dilator was measured by using an ad hoc device. Each modified dilator required significantly less force necessary for “tissue” insertion compared with the conical control (P < .05 for all designs). These data suggest that angled dilator tips may facilitate insertion and may contribute to new dilator and catheter tip design.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Vascular and Interventional Radiology - Volume 24, Issue 12, December 2013, Pages 1906-1909