کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
4240707 1283069 2007 4 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Significant Caval Penetration by the Celect Inferior Vena Cava Filter: Attributable to Filter Design?
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Significant Caval Penetration by the Celect Inferior Vena Cava Filter: Attributable to Filter Design?
چکیده انگلیسی
This report describes transmural penetration of the inferior vena cava (IVC) by a newly introduced IVC filter within 9 days of its placement. A computed tomographic study demonstrated filter penetration with one of the primary struts lodging in the uncinate process of the pancreas. Because of the close resemblance of this new filter to another filter that has not been associated with penetration injuries, the key differences between the two designs were examined and the determination was made that the new filter, unlike the older one, has unprotected primary struts. Only filters with an unprotected primary strut design have been associated with penetration injuries such as the one described in this case.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Vascular and Interventional Radiology - Volume 18, Issue 11, November 2007, Pages 1447-1450