کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5135794 1493441 2017 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Monitoring of vancomycin in human plasma via portable microchip electrophoresis with contactless conductivity detector and multi-stacking strategy
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Monitoring of vancomycin in human plasma via portable microchip electrophoresis with contactless conductivity detector and multi-stacking strategy
چکیده انگلیسی


- Multi-stacking technique based on FESI and MSS was developed in aqueous-based MCE-C4D.
- New approach was evaluated for the monitoring of vancomycin in human plasma.
- LOD of 1.2 μg/mL and SEF of 83-fold compared to typical gated injection method were obtained.

A portable microchip electrophoresis (MCE) coupled with on-chip contactless conductivity detection (C4D) system was evaluated for the determination of vancomycin in human plasma. In order to enhance the detection sensitivity, a new online multi-stacking preconcentration technique based on field-enhanced sample injection (FESI) and micelle-to-solvent stacking (MSS) was developed and implemented in MCE-C4D system equipped with a commercially available double T-junction glass chip. The cationic analytes from the two sample reservoirs were injected under FESI conditions and subsequently focused by MSS within the sample-loading channel. The proposed multi-stacking strategy was verified under a fluorescence microscope using Rhodamine 6G as the model analyte and a sensitivity enhancement factor (SEF) of up to 217 was achieved. The developed approach was subsequently implemented in the aqueous-based MCE, coupled to C4D in order to monitor the targeted antibiotic (in this case, vancomycin) present in human plasma samples. The multi-stacking and analysis time for vancomycin were 50 s and 250 s respectively, with SEF of approximately 83 when compared to typical gated injection. The detection limit of the method for vancomycin was 1.2 μg/mL, with intraday and interday repeatability RSDs of 2.6% and 4.3%, respectively. Recoveries in spiked human plasma were 99.0%-99.2%.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of Chromatography A - Volume 1485, 17 February 2017, Pages 142-146
نویسندگان
, , , ,