Etiology and Relevance of the Figure-of-Eight Artifact on Echocardiography after Percutaneous Left Atrial Appendage Closure with the Amplatzer Cardiac Plug
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Etiology and Relevance of the Figure-of-Eight Artifact on Echocardiography after Percutaneous Left Atrial Appendage Closure with the Amplatzer Cardiac Plug
چکیده انگلیسی
The figure-of-eight display of the ACP device during echocardiography is the result of the specific epitrochoid geometry of the device mesh and its interaction with ultrasound waves. It is important to recognize the figure-of-eight as being a normal imaging artifact of a correctly deployed device in the coronal imaging position on both transesophageal and transthoracic echocardiography. In the future, this could be used during follow-up to aid clinical practitioners in assessing device position and function.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of the American Society of Echocardiography - Volume 27, Issue 3, March 2014, Pages 323-328.e1