Dose comparison between TG-43-based calculations and radiochromic film measurements of the Freiburg flap applicator used for high-dose-rate brachytherapy treatments of skin lesions
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی تومور شناسی
پیش نمایش صفحه اول مقاله
Dose comparison between TG-43-based calculations and radiochromic film measurements of the Freiburg flap applicator used for high-dose-rate brachytherapy treatments of skin lesions
چکیده انگلیسی
For typical FF applicator loading sizes, the actual measured dose was 6% (±4%) lower at the skin level when compared with TG-43. Additional bolus above the FF was shown to decrease the dose difference. The consideration of change in clinical practice should be carefully investigated in light of clinical reference data.