کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
5699891 1410441 2007 6 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Comparison of 50 kV Facilities for Contact Radiotherapy
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی تومور شناسی
پیش نمایش صفحه اول مقاله
Comparison of 50 kV Facilities for Contact Radiotherapy
چکیده انگلیسی
The radiation characteristics of a short source to surface distance (SSD) contact therapy tube in clinical use at the Centre Antoine-Lacassagne, Nice and a long SSD unit at the Clatterbridge Centre for Oncology were compared. The output from the tube at Nice had a dose rate of approximately double that of the tube at Clatterbridge, whereas the tube at Clatterbridge had a slightly higher value of the half value layer. Depth dose measurements were made with GafChromic MD55 film and SSD corrected depth dose curves showed good agreement between centres. Profiles at 2 mm depth also showed comparable levels of flatness and uniformity.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Oncology - Volume 19, Issue 9, November 2007, Pages 655-660