Original ArticleA BMI-adjusted ultra-low-dose CT angiography protocol for the peripheral arteries-Image quality, diagnostic accuracy and radiation exposure
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی رادیولوژی و تصویربرداری
پیش نمایش صفحه اول مقاله
Original ArticleA BMI-adjusted ultra-low-dose CT angiography protocol for the peripheral arteries-Image quality, diagnostic accuracy and radiation exposure
چکیده انگلیسی

ObjectivesTo investigate radiation exposure, objective image quality, and the diagnostic accuracy of a BMI-adjusted ultra-low-dose CT angiography (CTA) protocol for the assessment of peripheral arterial disease (PAD), with digital subtraction angiography (DSA) as the standard of reference.MethodsIn this prospective, IRB-approved study, 40 PAD patients (30 male, mean age 72 years) underwent CTA on a dual-source CT scanner at 80 kV tube voltage. The reference amplitude for tube current modulation was personalized based on the body mass index (BMI) with 120 mAs for [BMI ≤ 25] or 150 mAs for [25 < BMI ≤ 30]. Iterative image reconstruction was applied. The presence of significant stenoses (>70%) was assessed by two readers independently and compared to subsequent DSA. Radiation exposure was assessed with the computed tomography dose index (CTDIvol) and the dosis-length product (DLP). Objective image quality was assessed via contrast- and signal-to-noise ratio (CNR and SNR) measurements. Radiation exposure and image quality were compared between the BMI groups and between the BMI-adjusted ultra-low-dose protocol and the low-dose institutional standard protocol (ISP).ResultsThe BMI-adjusted ultra-low-dose protocol reached high diagnostic accuracy values of 94% for Reader 1 and 93% for Reader 2. Moreover, in comparison to the ISP, it showed significantly (p < 0.001) lower CTDIvol (1.97 ± 0.55 mGy vs. 4.18 ± 0.62 mGy) and DLP (256 ± 81 mGy x cm vs. 544 ± 83 mGy x cm) but similar image quality (p = 0.37 for CNR). Furthermore, image quality was similar between BMI groups (p = 0.86 for CNR).ConclusionsA CT protocol that incorporates low kV settings with a personalized (BMI-adjusted) reference amplitude for tube current modulation and iterative reconstruction enables very low radiation exposure CTA, while maintaining good image quality and high diagnostic accuracy in the assessment of PAD.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: European Journal of Radiology - Volume 93, August 2017, Pages 149-156
نویسندگان
, , , , , , ,