Advanced virtual monoenergetic images: improving the contrast of dual-energy CT pulmonary angiography
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی تومور شناسی
پیش نمایش صفحه اول مقاله
Advanced virtual monoenergetic images: improving the contrast of dual-energy CT pulmonary angiography
چکیده انگلیسی


- Advanced monoenergetic image reconstruction from dual-energy CT results in less noise and higher CNR in pulmonary arteries.
- Subjective image quality is improved when using advanced monoenergetic reconstructions for dual-energy CT pulmonary angiography.
- Attenuation in the pulmonary artery circulation is higher in advanced monoenergetic image reconstructions.

AimTo investigate the value of advanced virtual monoenergetic image reconstruction (mono-plus) from dual-energy computed tomography (CT) for improving the contrast of CT pulmonary angiography (CTPA).Materials and methodsForty consecutive patients (25 women, mean 62.5 years, range 28-87 years) underwent 192-section dual-source CTPA with dual-energy CT (90/150 SnkVp) after the administration of 60 ml contrast media (300 mg iodine/ml). Conventional virtual monochromatic images at 60 keV and 17 mono-plus image datasets from 40-190 keV (in 10 keV steps) were reconstructed. Subjective image quality (artefacts, subjective noise) was rated. Attenuation was measured in the pulmonary trunk and in the right lower lobe pulmonary artery; noise was measured in the periscapular musculature. The signal-to-noise (SNR) and contrast-to-noise ratios (CNR) were calculated for each patient and dataset. Comparisons between monochromatic images and mono-plus images were performed by repeated measures analysis of variance (ANOVA) with post-hoc Bonferroni correction.ResultsInterreader agreement was good to excellent for subjective image quality (ICC: 0.616-0.889). As compared to conventional 60 keV images, artefacts occurred less (p=0.001) and subjective noise was rated lower (p<0.001) in mono-plus 40 keV images. Noise was lower (p<0.001), and the SNR and CNR in the pulmonary trunk and right lower lobe pulmonary artery were higher (both, p<0.001) in mono-plus 40 keV images compared to conventional monoenergetic 60 keV images. Transient interruption of contrast (TIC) was found in 14/40 (35%) of patients, with subjective contrast being similar 8/40 (20%) or higher 32/40 (80%) in mono-plus 40 keV as compared to conventional monoenergetic 60 keV images.ConclusionsCompared to conventional virtual monoenergetic imaging, mono-plus images at 40 keV improve the contrast of dual-energy CTPA.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Clinical Radiology - Volume 70, Issue 11, November 2015, Pages 1244-1251
نویسندگان
, , , , , ,