کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
8249006 1533350 2018 9 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Intercomparison of 99mTc, 18F and 111In activity measurements with radionuclide calibrators in Belgian hospitals
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم تشعشع
پیش نمایش صفحه اول مقاله
Intercomparison of 99mTc, 18F and 111In activity measurements with radionuclide calibrators in Belgian hospitals
چکیده انگلیسی
The overall results of the intercomparison were satisfactory for 99mTc and 18F, since most radionuclide calibrators (>70%) were accurate within ±5% of the reference value. Nevertheless, some devices underestimated the activity by 10-20%. Conversely, 111In measurements were strongly affected by source geometry effects and this had a negative impact on the accuracy of the measurements, in particular for the syringe sample. Large overestimations (up to 72%) were observed, even when taking into account the corrections and uncertainties supplied by the manufacturers for container effects. The results of this exercise encourage the hospitals to perform corrective actions to improve the calibration of their devices where needed.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Physica Medica - Volume 45, January 2018, Pages 134-142
نویسندگان
, , , , , , ,