CT-ultrasound fusion prostate brachytherapy: A dynamic dosimetry feedback and improvement method. A report of 54 consecutive cases
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی تومور شناسی
پیش نمایش صفحه اول مقاله
CT-ultrasound fusion prostate brachytherapy: A dynamic dosimetry feedback and improvement method. A report of 54 consecutive cases
چکیده انگلیسی
CT-US fusion prostate brachytherapy represents a dynamic dosimetry feedback and remediation method that consistently produced high prostate V100 and D90 values with acceptably low urethra D50 and D10 values in our study. The degree of prostate V100 and D90 dosimetry improvement created by remedial source placement effectively matched the degree of initial dosimetry deficiency. This method produced a high level of correlation between the preplan and final prostate D90 values.