کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
9971191 1577922 2005 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Doppler echocardiography of 119 normal-functioning St Jude Medical mitral valve prostheses: A comprehensive assessment including time-velocity integral ratio and prosthesis performance index*
موضوعات مرتبط
علوم پزشکی و سلامت پزشکی و دندانپزشکی کاردیولوژی و پزشکی قلب و عروق
پیش نمایش صفحه اول مقاله
Doppler echocardiography of 119 normal-functioning St Jude Medical mitral valve prostheses: A comprehensive assessment including time-velocity integral ratio and prosthesis performance index*
چکیده انگلیسی
The purpose of this study was to provide, in a large number of patients, comprehensive Doppler echocardiographic assessment of normal St Jude Medical mitral valve prosthesis function using Doppler-derived hemodynamic variables, including the mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio and prosthesis performance index. The pressure half-time was less than 130 milliseconds in all patients, and all but one patient had either a peak early mitral diastolic velocity of 2 m/s or less or a mitral valve prosthesis-to-left ventricular outflow tract time-velocity integral ratio of less than 2.2. There was a significant (P < .001) negative correlation between the prosthesis performance index and prosthesis size. This negative correlation suggests that there is more efficient use of the in vitro geometric orifice area with smaller prostheses.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Journal of the American Society of Echocardiography - Volume 18, Issue 3, March 2005, Pages 252-256
نویسندگان
, , , , , , ,