کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
10434932 910788 2016 8 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Optimal calibration of instrumented treadmills using an instrumented pole
ترجمه فارسی عنوان
کالیبراسیون بهینه از تردمیل های ابزار دقیق با استفاده از یک قطب ابزار
موضوعات مرتبط
مهندسی و علوم پایه سایر رشته های مهندسی مهندسی پزشکی
چکیده انگلیسی
Calibration of instrumented treadmills is imperative for accurate measurement of ground reaction forces and center of pressure (COP). A protocol using an instrumented pole has been shown to considerably increase force and COP accuracy. This study examined how this protocol can be further optimized to maximize accuracy, by varying the measurement time and number of spots, using nonlinear approaches to calculate the calibration matrix and by correcting for potential inhomogeneity in the distribution of COP errors across the treadmill's surface. The accuracy increased with addition of spots and correction for the inhomogeneous distribution across the belt surface, decreased with reduction of measurement time, and did not improve by including nonlinear terms. Most of these methods improved the overall accuracy only to a limited extent, suggesting that the maximal accuracy is approached given the treadmill's inherent mechanical limitations. However, both correction for position dependence of the accuracy as well as its optimization within the walking area are found to be valuable additions to the standard calibration process.
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Medical Engineering & Physics - Volume 38, Issue 8, August 2016, Pages 785-792
نویسندگان
, , , ,