A modified aerospray apparatus was used to prepare a thin layer sample of matrix and analyte for quantitative analysis by MALDI-TOFMS. The apparatus consists of a set of coaxial tubing; the liquid sample is forced by a syringe pump through the inner capillary and it is nebulized by a flow of gas through the outer capillary. The small droplets of sample exiting the device are deposited onto a rotating plate, which serves as the sample surface for a time-of-flight mass spectrometer. An optimization was carried out after initial experiments with the device resulted in poorer than expected reproducibility of analyte signal. A two-level plus center point factorial experiment was performed investigating several factors, including the inner capillary internal diameter, gas pressure, liquid flow, spray distance, and time. After optimization the within-sample reproducibility of the analyte signal improved 3-fold, while the sample-to-sample reproducibility improved 4.5-fold.
Journal: Analytica Chimica Acta - Volume 658, Issue 1, 18 January 2010, Pages 49–55