کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1167943 1491140 2010 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Selective determination of 2,4-xylenol by gas chromatography/supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Selective determination of 2,4-xylenol by gas chromatography/supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometry
چکیده انگلیسی

Gas chromatography/supersonic jet/resonance-enhanced multiphoton ionization/time-of-flight mass spectrometry (GC/SSJ/REMPI/TOF-MS) was employed for isomer-selective determination of 2,4-xylenol in river and seawater samples. The sample containing 2,4-xylenol was measured using argon, rather than helium, as the GC carrier gas to cool the analyte molecule sufficiently. The instrumental detection limit (IDL) achieved at a flow rate of 1 mL min−1 was 14 pg. Although this value was comparable to the value (ca. 10 pg) obtained by gas chromatography/electron impact/quadrupole mass spectrometry (GC/EI/QMS). When the flow rate was increased to 8 mL min−1, interference from the 2,5-xylenol isomer was completely suppressed. The IDL was degraded to 83 or 160 pg at a flow rate of 5 or 8 mL min−1, respectively. The recovery of 2,4-xylenol from the river and the seawater samples was 85 and 93%, respectively. The time for analysis was only 10 min per one sample in GC/SSJ/REMPI/TOF-MS. These results suggest that GC/SSJ/REMPI/TOF-MS is useful for the selective measurement of 2,4-xylenol, which has been designated a Class I chemical substance in the Pollutant Release and Transfer Register (PRTR).

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Analytica Chimica Acta - Volume 682, Issues 1–2, 3 December 2010, Pages 72–76
نویسندگان
, , , , ,