Determination of Platinum in Human Plasma by Microchip-based Nanoflow Injection and Inductively Coupled Plasma Mass Spectrometry
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Determination of Platinum in Human Plasma by Microchip-based Nanoflow Injection and Inductively Coupled Plasma Mass Spectrometry
چکیده انگلیسی

A microfluidic chip-based nanoflow injection system was set up for inductively coupled plasma mass spectrometry (ICP-MS). The sample plug loaded into the sampling channel of the microchip was driven into ICP-MS, as soon as the multi-functional valve was changed from the load position to the inject position. The sample plug shape depends on the sampling channel. 40–200 nL sample was introduced into ICP-MS with 5–25 mm sampling channel. The proposed microchip-based nanoflow injection sampling system offers many advantages such as low sample consumption, satisfactory precisions (less than 3.0%), low sampling dead volume (nearly zero) and easy fabrication etc. With 200-nL sampling volume and a flow rate of 20 μL min−1 for the carrier, a peak profile with the highest peak height within the shortest time was obtained. The absolute detection limit of the chip-based nanoflow injection sampling system for 195Pt was 2.54 fg under the optimized conditions, which was improved by a factor of 3200 in comparison with the conventional sampling system. A sample throughput of 48 h−1 was obtained with the nanoflow injection sampling system. Relative standard deviations of peak heights of 10 replicate trials of 10 μg L−1 standard solution and the human plasma sample S1 were 1.71% and 2.65%, respectively. The contents of platinum in six human plasma samples detected by the proposed method agreed well with those by the conventional sampling system, and the recoveries of six plasma samples ranged from 94.3% to 103.0%. These results indicate the good accuracy of the present method.

A microchip-based nanoflow injection system was used for the analysis of limited biological samples by inductively coupled plasma mass spectrometry, which offers outstanding advantages including low sample consumption (50–200 nL), satisfactory precisions, low sampling dead volume (nearly zero) and easy fabrication.Figure optionsDownload as PowerPoint slide

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Chinese Journal of Analytical Chemistry - Volume 41, Issue 3, March 2013, Pages 349–353
نویسندگان
, , , , ,