کد مقاله | کد نشریه | سال انتشار | مقاله انگلیسی | نسخه تمام متن |
---|---|---|---|---|
1224154 | 967912 | 2008 | 6 صفحه PDF | دانلود رایگان |

The present study reports the optimization and validation of a dissolution test for selegiline·HCl tablets using a new high-performance liquid chromatographic (HPLC) method. Rapid separation of the analyte from sample matrix was achieved in less than 60 s using a Cromolith® RP-18e monolithic column using UV detection at 220 nm. Thorough validation of the assay based on pre-defined criteria included linearity, LOD/LOQ, accuracy, precision, selectivity and ruggedness. The dissolution test was optimized in terms of dissolution medium, basket (type I)/paddle (type II) agitation and rotation speed. Its ruggedness was also validated. The presented analytical and dissolution procedures are currently being applied in the quality and stability control of Cosmopril® tablets (5 mg/tablet selegiline·HCl, Cosmopharm Ltd., Korinthos, Greece).
Journal: Journal of Pharmaceutical and Biomedical Analysis - Volume 46, Issue 4, 13 March 2008, Pages 670–675