کد مقاله کد نشریه سال انتشار مقاله انگلیسی نسخه تمام متن
1246827 969773 2009 5 صفحه PDF دانلود رایگان
عنوان انگلیسی مقاله ISI
Development of urease based amperometric biosensors for the inhibitive determination of Hg (II)
موضوعات مرتبط
مهندسی و علوم پایه شیمی شیمی آنالیزی یا شیمی تجزیه
پیش نمایش صفحه اول مقاله
Development of urease based amperometric biosensors for the inhibitive determination of Hg (II)
چکیده انگلیسی

Enzymatic amperometric procedures for measurement of Hg (II), based on the inhibitive action of this metal on urease enzyme activity, were developed. Screen-printed carbon electrodes (SPCEs) and gold nanoparticles modified screen-printed carbon electrodes (AuNPs/SPCEs) were used as supports for the cross-linking inmobilization of the enzyme urease. The amperometric response of urea was affected by the presence of Hg (II) ions which caused a decreasing in the current intensity. The optimum working conditions were found using experimental design methodology. Under these conditions, repeatability and reproducibility for both types of biosensors were determined, reaching values below 6% in terms of residual standard deviation. The detection limit obtained for Hg (II) was 4.2 × 10−6 M for urease/SPCE biosensor and 5.6 × 10−8 M for urease/AuNPs/SPCE biosensor. Analysis of the possible effect of the presence of foreign ions in the solution was performed. The method was applied to determine levels of Hg (II) in spiked human plasma samples.

ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Talanta - Volume 79, Issue 5, 15 October 2009, Pages 1306–1310
نویسندگان
, , , ,