Feasibility of in vivo Transesophageal Cardiac Ablation Using a Phased Ultrasound Array
موضوعات مرتبط
مهندسی و علوم پایه فیزیک و نجوم آکوستیک و فرا صوت
پیش نمایش صفحه اول مقاله
Feasibility of in vivo Transesophageal Cardiac Ablation Using a Phased Ultrasound Array
چکیده انگلیسی
Over 2.2 million Americans suffer from atrial fibrillation making it one of the most common arrhythmias. Cardiac ablation has shown a high rate of success in treating paroxysmal atrial fibrillation. Prevailing modalities for this treatment are catheter based radio-frequency ablation or surgery. However, there is measurable morbidity and significant costs and time associated with these invasive procedures. Due to these issues, developing a method that is less invasive to treat atrial fibrillation is needed. In the development of such a device, a transesophageal ultrasound applicator for cardiac ablation was designed, constructed and evaluated. A goal of this research was to create lesions in myocardial tissue using a phased array. Based on multiple factors from array simulations, transesophageal imaging devices and throat anatomy, a phased ultrasound transducer that can be inserted into the esophagus was designed and tested. In this research, a two-dimensional sparse phased array with the aperture size of 20.7 mm × 10.2 mm with flat tapered elements as a transesophageal ultrasound applicator was fabricated and evaluated with in vivo experiments. Five pigs were anesthetized; the array was passed through the esophagus and positioned over the heart. The array was operated for 8 ∼ 15 min at 1.6 MHz with the acoustic intensity of 150 ∼ 300 W/cm2 resulting in both single and multiple lesions on atrial and ventricular myocardium. The average size of lesions was 5.1 ± 2.1 mm in diameter and 7.8 ± 2.5 mm in length. Based on the experimental results, the array delivered sufficient power to the focal point to produce ablation while not grossly damaging nearby tissue outside the target area. These results demonstrate a potential application of the ultrasound applicator to transesophageal cardiac surgery in atrial fibrillation treatment. (E-mail: Jrw140@psu.edu)
ناشر
Database: Elsevier - ScienceDirect (ساینس دایرکت)
Journal: Ultrasound in Medicine & Biology - Volume 36, Issue 5, May 2010, Pages 752-760
نویسندگان
, , , , ,